SARS-CoV-2/influenza A /influenza B
Nama produk
HWTS-RT148-SARS-CoV-2/influenza A /influenza B Kit Pengesanan Gabungan Asid Nukleik (PCR Pendarfluor)
Saluran
Nama Saluran | Campuran PCR 1 | Campuran PCR 2 |
Saluran FAM | gen ORF1ab | IVA |
Saluran VIC/HEX | Kawalan dalaman | Kawalan dalaman |
Saluran CY5 | gen N | / |
Saluran ROX | gen E | IVB |
Parameter Teknikal
Penyimpanan | -18 ℃ |
Jangka hayat | 12 bulan |
Jenis Spesimen | swab nasofaring dan swab oropharyngeal |
Sasaran | SARS-CoV-2 tiga sasaran ( gen Orf1ab, N dan E )/influenza A /influenza B |
Ct | ≤38 |
CV | ≤10.0% |
LoD | SARS-CoV-2:300 Salinan/mL virus influenza A:500 Salinan/mL virus influenza B:500 Salinan/mL |
Kekhususan | a) Keputusan ujian silang menunjukkan bahawa kit itu serasi dengan coronavirus manusia SARSr- CoV, MERSr-CoV, HcoV-OC43, HcoV-229E, HcoV-HKU1, HCoV-NL63, virus pernafasan pernafasan A dan B, virus parainfluenza 1, 2 dan 3, virus rhino, adeno, B dan 3. 5, 7 dan 55, metapneumovirus manusia, enterovirus A, B, C dan D, virus pulmonari sitoplasma manusia, virus EB, virus campak Sitomegalovirus manusia, rotavirus, norovirus, virus beguk, virus varicella zoster, Mycoplasma pneumoniae, Chlamydia pneumoniae, Legionella, pertusis, Staphilus a. Streptococcus pneumoniae, Streptococcus pyogenes, Klebsiella pneumoniae, Mycobacterium tuberculosis, Aspergillus fumigatus, Candida albicans, Candida glabrata Tiada tindak balas silang antara Pneumocystis yersini dan Cryptococcus neoformans. b) Keupayaan anti gangguan: pilih mucin (60mg/mL), 10% (V/V) darah manusia, diphenylephrine (2mg/mL), hydroxymethylzoline (2mg/mL), natrium klorida (mengandungi bahan pengawet) (20mg/mL), beclomethasone (20mg/mL), dexamethasone (20mg/mL), dexamethasone (20mg/mL), flunisonemL (20mg/mL), flunisonemL. triamcinolone acetonide (2mg/mL), budesonide (2mg/mL), mometasone (2mg/mL), fluticasone (2mg/mL), histamine hydrochloride (5mg/mL), α-Interferon (800IU/mL), zanamivir (20mg/mL), ribavirin (10mg/mL), ribavirin (10mg/mL) (1mg/mL), lopinavir (500mg/mL), ritonavir (60mg/mL), mupirocin (20mg/mL), azithromycin (1mg/mL), ceprotene (40μg/mL) Meropenem (200mg/mL), levofloxacin (10μg/mL) dan tobramycin (10μg/mL) dan tobramycin (10μg/mL). Keputusan menunjukkan bahawa bahan pengganggu pada kepekatan di atas tidak mempunyai tindak balas gangguan terhadap hasil pengesanan patogen. |
Instrumen Berkenaan | Applied Biosystems 7500 Sistem PCR Masa Nyata Applied Biosystems 7500 Sistem PCR Masa Nyata Pantas SLAN ®-96P Sistem PCR Masa Nyata Sistem PCR Masa Nyata QuantStudio™ 5 LightCycler®480 Sistem PCR Masa Nyata Sistem Pengesanan PCR Masa Nyata LineGene 9600 Plus MA-6000 Pesikal Terma Kuantitatif Masa Nyata Sistem PCR Masa Nyata BioRad CFX96 Sistem PCR Masa Nyata BioRad CFX Opus 96 |
Jumlah Penyelesaian PCR